In this study, human fingernails were used to estimate radiation dose by electron paramagnetic resonance (EPR) spectroscopy. The main limitation was the presence of mechanically induced EPR signals generated during nail clipping and sample preparation. Because these signals overlap with the radiation-induced signal, the absorbed dose can be overestimated. Several treatments were therefore applied to reduce the mechanical signal. The results showed that mechanically induced and radiation-induced signals have different microwave power saturation behavior. The mechanical signal faded with time and reached a constant value after about 20 h. Heat treatment at 70°C for 3 h reduced the mechanical signal to the same extent as 20 h at ambient temperature. Among chemical treatments, dithiothreitol produced the lowest mechanical signal. Dose-response curves for nails treated with dithiothreitol for 30 min were more linear and precise than those treated for 5 min. Therefore, for unknown absorbed-dose estimation in fingernail samples using a calibration curve, treatment with dithiothreitol for 30 min is recommended to reduce uncertainty.